YL96
Laser Photocoagulator
Treatment Beam
Type: Frequency-doubled Nd:GdVO4 Laser
Wavelength: Green: 532 nm
Power output: Adaptor output:
50mW~1,200mW adjustable by step;
50mW~100mW, 5mW step;
100mW~300mW, 10mW step;
300mW~800mW, 20mW step;
800mW~1,000mW, 50mW step;
≥1000mW, 100mW step;
Allowable tolerance between actual output power and pre-set value is within ±15%.
Pulse duration: 0.01s ~ 3s ± 5%
Pulse interval: 0.05s ~ 1s ± 5%; or Single Pulse (SP)
Duty cycle1: When output power is within 50mW ~ 500mW , duty cycle is not limited;
When output power is within 500mW ~ 1000mW , duty cycle is limited to 60%;
When output power is above1000mW , duty cycle is limited to 50%.
Pulse Counter: 0~9,999
Laser beam spot size: 50µm~500µm, continuously adjustable, the allowable tolerance of scale is within ±15%; the allowable tolerance of minimum scale (50µm) is within ±15µm.
Beam divergence angle: Adaptor output, 13.8°± 20%
Repeatability of laser power: RP ≤±10 %
Aiming Beam
Type: Semi-conductor Laser
Wavelength: Red: 630nm~680nm
Output Power: ≤1mW, adjustable by level
Cooling System
Primary Cooling System: Intelligent TEC (thermal-electric-coupling) semi-conductor cooling; Secondary Cooling System: Air-Cooled.
|